Specifications
|
9
ELx800 Operator's Manual
The following specifications apply to a 384-well plate. The minimum volume of
fluid in a well to obtain the specified performance is 80 microliters. The
specifications apply to the dual wavelength mode of read only.
Absorbance Measurement
Range:
0.000 to 3.000 OD
Normal Read Mode:
Dual wavelength (4 minutes, 45 seconds)
Accuracy:
± 2.0% ± 0.020 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity:
± 2.5% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD):
± 1.5% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm
Rapid Read Mode:
Dual wavelength (3 minutes, 25 seconds)
Accuracy:
± 2.5% ± 0.020 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity:
± 2.5% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD):
± 2.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...