![Biotek EL800 Operator'S Manual Download Page 178](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074178.webp)
152
|
Appendix A:
Computer Control
BioTek Instruments, Inc.
Controlling the Reader With KC4™
Before installing KC4 software, verify that your computer meets the
minimum system requirements specified in KC4’s User’s Guide or Help
system.
Setting Up KC4
The ELx800™ can be operated via BioTek's KC4 software installed on a host computer.
The following instructions
briefly
show you how to set up KC4 for operation of the
reader. Refer to KC4’s User’s Guide or Help system for more detailed instructions.
1.
Turn off the computer and the reader. Connect the appropriate serial cable
(PN 75053) between the two machines.
2.
Turn on both machines.
3.
Install KC4 on the computer’s hard drive, and register the software with
BioTek.
4.
Once installed, start KC4.
5.
Select
System|Readers
.
6.
Scroll through the list of
Available Readers
and select the appropriate
ELx800 reader model. Click the
Port
button (and subsequent
Setup
button), to
define the following communications parameters:
Port:
Select the COM port used for the RS-232
serial cable connection
Transmission Speed:
9600
(default),
1200
, or
2400
(must match
the baud rate on the reader)
Data Bits:
8
Parity:
No
Stop Bits:
2
7.
Click the
Current Reader
button to attempt to establish communication with
the reader, using the currently defined communication parameters.
8.
If the test passes, click
OK
to save the settings and close the dialog box. If the
test fails, KC4 will provide appropriate instructions for resolving any
problems. See also the Problems section on the following page.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...