Chapter 6: Troubleshooting and Error Codes
BioTek Instruments, Inc.
General Errors (Continued)
0900
Read time dark value out of range
•
The dark current value taken during the current read is significantly
different from the same reading taken during the power-up self-check.
Probable Cause:
•
The measurement electronics background noise has changed since the last
power-up self-check.
Ambient light may have increased since power-up.
0A01-
0A05
Read time Air Blank out of range
The blank (full signal) reading taken during the current read has changed
significantly from the same reading taken during the power-up self-check.
Probable Cause:
•
Lamp is unstable. Reboot system and/or install new lamp. Turn unit
OFF
when not in use to conserve lamp life.
0B00 Invalid
assay
An assay number that is not programmed was selected.
0C00
Printer timed out
The time allotted for the instrument to make a valid connection to a printer has
expired.
Probable Causes:
•
Printer not connected, on line, or powered up.
•
Printer parallel port may not be correctly selected.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...