![Biotek EL800 Operator'S Manual Download Page 85](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074085.webp)
Define (Method, Map, Formula and Curve)
|
59
ELx800 Operator’s Manual
Blank Location
The
BLANK LOCATION
screen allows you to define where the blank well or
wells
occur on the microplate. This screen only appears if
manual mapping
was selected.
E N T E R T H E L O C A T I O N O F
B
L
A
N
K
# 1 :
A
1 2
•
Use the
numeric
and
alpha
keys to enter a Blank Location, based upon
the plate geometry.
•
Use the arrow keys to move the cursor to the next or previous editable
field. The cursor is positioned beneath the first editable field.
•
Press
ENTER
to continue.
Number of Standards
This option allows you to enter the number of standard
groups
that will be
used in the assay. Any previously defined value will be displayed on the
screen.
Note:
If the number of standards is altered, the number of
replicates for the standard automatically reverts
to 1.
E N T E R N U M B E R O F
S
T
A
N
D
A
R
D S :
0 2
•
Use the
numeric
keys to enter the
NUMBER OF STANDARDS
. The valid
range depends on the selected curve fit method. The maximum number
of standards is 12. The minimum is 4 for 4-P fit, cubic, cubic spline, and
logit-log; 3 for quadratic; and 2 for linear and point-to-point.
•
Press
CLEAR
to clear the value on the display.
•
Press
ENTER
to continue.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...