Perform a 384-Well Plate Read in KCjunior™
|
167
ELx800 Operator’s Manual
9.
Select a
Read Method Type
.
10.
Select the wavelength(s) at which the plate will be read.
11.
Set the
Plate Geometry
to
16 x 24
.
12.
Define other reading parameters, as necessary. Click the
Help
button for
assistance.
13.
When you have finished defining the parameters, click
Validate Read
Method
if you would like KCjunior™ to verify the defined parameters. If any
parameters are invalid (or incomplete), follow the instructions provided by
KCjunior for correcting the parameters.
Figure 37: Example Read Method parameters
14.
Place the 384-well plate on the plate carrier, then click
OK
to read the plate.
Click
OK
when prompted to “Place the plate on the carrier.”
The plate will be read and then the raw data results will display in
KCjunior.
To analyze or manipulate the results,
Protocol
parameters should be
defined. Refer to KCjunior's Help system or User's Guide for
instructions.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...