![Biotek EL800 Operator'S Manual Download Page 141](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074141.webp)
Liquid Testing
|
115
ELx800 Operator's Manual
Stock Solution Formulation
The stock solution for Liquid Tests No. 1 and No. 2 may be formulated from the
ingredients listed below (Solution A), or by diluting a dye solution available from
BioTek (Solution B).
Solution A
Required Materials:
•
Deionized (DI) water
•
FD&C Yellow No. 5 dye powder (typically 90% pure)
•
Tween
®
20 (polyoxyethylene (20) sorbitan monolaurate)
or
BioTek
wetting agent, PN 7773002
•
Precision balance with readability of 0.001 g
•
Weigh boat
•
1-liter volumetric flask
Preparation of Stock Solution:
1
Weigh out 0.092 gram of FD&C No. 5 yellow dye powder into a weigh
boat.
2
Rinse the contents into a 1-liter volumetric flask.
3
Add 0.5 ml of Tween 20, or 5 ml of BioTek’s wetting agent.
4
Make up to 1 liter with DI water, cap, and shake well.
This should create a solution with an absorbance of about 2.000 when
using 200 µl in a flat-bottom microwell. The OD value will be
proportional to the volume in the well and the amount of FD&C No. 5
dye used. You can use a larger or smaller well volume, or add more
dye or water to adjust the solution. Note that too small a well volume
may result in increased pipetting-related errors.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...