Liquid Testing
|
125
ELx800 Operator's Manual
Procedure
1.
Check the absorbance of a sample of buffer solution A or B at 340 nm on the
microplate reader. This solution, which will be referred to as the
100%
Test Solution
, will have an optical density (absorbance) of approximately
0.700 to 1.000. This value is not critical, but it should be within this
absorbance range.
If low, adjust up by adding
D
-NADH powder until the solution is at least at
the lower end of this range. Do not adjust if slightly high.
2.
Carefully prepare a
75% Test Solution
by diluting 15 ml of the 100% Test
Solution:
If using the Sigma PBS Solution, use 5 ml as the diluent.
If using the 10x Concentrate PBS Solution, mix one part of the
concentrate with nine parts of DI water. Use 5 ml of this solution as the
diluent.
3.
Carefully prepare a
50% Test Solution
by diluting 10 ml of the 100% Test
Solution:
If using the Sigma PBS Solution, use 10 ml as the diluent.
If using the 10x Concentrate PBS Solution, mix one part of the
concentrate with nine parts of DI water. Use 10 ml of this solution as the
diluent.
4.
Pipette the three solutions into a new 96-well microplate:
150 µl of the 100% Test Solution into all wells of columns 1 and 2
150 µl of the 75% Test Solution into all wells of columns 3 and 4
150 µl of the 50% Test Solution into all wells of column 5 and 6
After pipetting the diluted test solution into the microplate and
before
reading the plate, we strongly recommend
shaking
the plate for four
minutes. This will allow any air bubbles in the solution to settle and the
meniscus to stabilize. If a plate shaker is not available, wait 20
minutes after pipetting the diluted test solution before reading the
plate.
5.
Read the microplate five times using Normal Mode, single wavelength at
340 nm, no blanking (or blank on air).
6.
Print the five sets of raw data or export it to an Excel spreadsheet using
Gen5™, KC4™, or KCjunior™.
If you are exporting the data to an Excel spreadsheet, perform the
calculations described on the following page, and keep the spreadsheet
for future tests.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...