Liquid Testing
|
121
ELx800 Operator's Manual
Select
Tools|Data Analysis|Regression
. Identify column X as the
“Input X Range” and column Y as the “Input Y Range” and then click
OK to perform the analysis.
Note:
If the Data Analysis command is not available on the Tools
menu, you may need to install the Analysis ToolPak in Microsoft
®
Excel. Consult Excel’s help system for assistance.
Expected Results:
Since it is somewhat difficult to achieve high pipetting accuracy when
conducting linear dilutions, an R-Square value greater than or equal to
0.990 is considered adequate.
Repeatability Test
1
Calculate the mean and standard deviation for the five readings taken
above at each concentration. Only one data set needs to be analyzed for
each concentration. The well that shows the most variation for each
concentration is selected for data reduction.
2
For each mean below 2.000 OD, calculate the allowed deviation using
the repeatability specification for a 96-well format of ± 0.5% ± 0.005 OD
from 0.000 to 2.000 OD @ 405 nm.
3
The standard deviation for each set of readings should be less than the
allowed deviation.
For example:
Absorbance readings of 1.950, 1.948, 1.955, 1.952, and 1.950 will result in
a mean of 1.951, and a standard deviation of 0.0026. The mean (1.951)
multiplied by 0.5% (1.951 * 0.005) = 0.0098, which, when added to the
0.005 (0.0098 + 0.005) = 0.0148 OD, which is the allowable deviation.
Since the standard deviation is less than this value, the reader meets the
test criteria.
Repeatability Specification:
For comparison in this test, the following repeatability spec is applied,
using Normal mode, 96-well microplate:
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @405 nm
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...