Chapter 4: Instrument Qualification
BioTek Instruments, Inc.
Requirements
To run the Absorbance Plate Test, you need BioTek's Absorbance Test Plate
(PN 9000547 or PN 7260522), with its accompanying Standards Certificate, shown
in
Figure 18
.
Figure 19
shows a sample Test Plate Analysis Report.
This test plate can be used for testing the reproducibility, linearity, and alignment of
your BioTek autoreader. The following calibration data has been recorded by a
N.I.S.T. traceable spectrophotometer.
WAVELENGTH (nm)
Well 405nm 450nm 490nm 550nm 620nm 630nm 690nm 750nm
C1 0.147 0.140 0.135 0.130 0.136 0.136 0.127 0.134
E2 0.618 0.575 0.574 0.568 0.573 0.568 0.485 0.434
G3 1.133 1.052 1.051 1.040 1.050 1.040 0.881 0.783
H6 1.701 1.578 1.577 1.560 1.575 1.560 1.323 1.179
F5 2.279 2.024 1.976 1.956 1.893 1.865 1.537 1.272
D4 2.945 2.604 2.545 2.513 2.437 2.400 1.972 1.632
Set # 2453
Serial # 161259
Figure 17: Sample Standards Certificate for the Absorbance Test Plate
Alignment Results:
WELLS B2 B12 G1 G11
DATA 0.000 0.000 0.000 0.000
TOLERANCE 0.015 0.015 0.015 0.015
RESULT PASS PASS PASS PASS
Wavelength = 405 nm
Accuracy Results:
WELLS C1 E2 G3 H6 F5 D4
STANDARD 0.147 0.618 1.133 1.701 2.279 2.945
MIN. VALUE 0.124 0.586 1.090 1.647 #N/A #N/A
MAX. VALUE 0.170 0.650 1.176 1.755 #N/A #N/A
DATA 0.148 0.616 1.129 1.695 2.243 2.850
RESULT PASS PASS PASS PASS #N/A #N/A
Repeatability Results:
WELLS C1 E2 G3 H6 F5 D4
READ1 0.148 0.616 1.129 1.695 2.243 2.850
MIN. VALUE 0.137 0.600 1.108 1.668 #N/A #N/A
MAX. VALUE 0.159 0.632 1.150 1.722 #N/A #N/A
READ2 0.149 0.616 1.130 1.695 2.242 2.853
RESULT PASS PASS PASS PASS #N/A #N/A
Figure 18: Sample Test Plate Analysis Report
Note:
The Test Plate Analysis report varies slightly depending on how it
was run (via the keypad, Gen5™, KC4™, or KCjunior™).
Important!
The ELx800’s Absorbance Plate Test tests the accuracy and
repeatability specifications from 0.000 to 2.000 OD only. In the example
above, the Analysis report displays the OD values read in well positions
D04 and F05 but does not indicate PASS or FAIL, because the values are
higher than 2.000 OD and therefore are not within the test range.
S A M P L E
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...