Controlling the Reader With KCjunior™
|
155
ELx800 Operator's Manual
6.
Click the
Test Communications
button to attempt to establish
communications with the reader, using the currently defined communication
parameters. If a
Serial Write Error
dialog box is displayed, an incorrect Com
Port may have been selected. Select a different port and then repeat this step.
7.
If the test passes, click
OK
to save the settings and close the dialog box. If the
test fails, follow the directions provided by KCjunior, then click
Test
Communications
again.
Problems
If KCjunior fails to communicate with the reader, and displays a serial
communications error, check the cable plug-in location to make sure it matches the
setup choices and is not a Null cable. If this is suspected, add another Null and try
again.
Getting Started With KCjunior™
The following instructions
briefly
describe how to read a plate using KCjunior.
Refer to KCjunior's Help system and User's Guide early and often to learn how to
create protocols, assign well identifiers, read plates, print reports, and more.
To read a plate using KCjunior:
1.
Click
Read Plate
from KCjunior's main screen. The ‘Read Plate Dialog’
will appear.
2.
If desired, enter a Results ID and a Plate Description, and then click
Read
Plate.
The ‘Protocol Definition’ dialog will appear.
3.
Select a
Read Method Type.
4.
Define the wavelength(s) at which the plate will be read.
5.
Select a
Plate Geometry
from the drop-down list.
6.
Define other reading parameters as necessary. Click the Help button for
assistance.
7.
When complete, click OK to return to the Read Plate Dialog. If desired,
enter a
Plate ID.
8.
Place the plate on the carrier, then click
OK
to start the plate read.
The plate will be read and then the raw data results will display in
KCjunior.
To analyze, manipulate, or print results,
protocol
parameters should be
defined. Refer to KCjunior's Help system or User's Guide for
instructions.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...