104
|
Chapter 4: Instrument Qualification
BioTek Instruments, Inc.
Qualification Procedures
You may use the tests outlined in this section to confirm initial and ongoing performance
of the ELx800™
.
Set up the ELx800 according to the instructions in
Chapter 2,
Installation
.
Confirm that the instrument powers up and communicates with peripherals
for the Installation Qualification. After Installation Qualification, conduct the Operational
Qualification Tests.
Note:
An Installation Qualification/Operational Qualification/Performance
Qualification (
IQ/OQ/PQ
) package (PN 7330538) for the ELx800 is available
for purchase. Contact your local dealer for more information.
Your ELx800
reader was fully tested at BioTek before shipment and should operate
properly upon initial setup. If you suspect that problems may have occurred during
shipment, if you reshipped the instrument, or if regulatory requirements dictate that
Performance Qualification Testing is necessary, you should perform the following tests.
After the initial confirmation of operation, you should perform the Absorbance Plate Test
monthly and Liquid Testing semiannually.
•
The
System Test
(described on the next page) confirms that the light levels at all
installed filter wavelengths and electronic noise meet factory acceptance criteria.
•
The
Checksum Test
(page 107) compares the onboard software with internally
recorded checksum values to ensure that no corruption has occurred.
•
The
Absorbance Plate Test
(page 108) uses BioTek’s Absorbance Test Plate to
confirm the mechanical alignment, optical accuracy/linearity, and repeatability of
the instrument.
•
Liquid Testing
(page 114) uses liquid solutions in a microplate to confirm
mechanical alignment, optical accuracy/linearity, and repeatability of the
instrument.
Set the ELx800 to
Normal Read
Mode
when running
all verification procedures (see
Selecting Read
Options
in
Chapter 3
).
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...