Chapter 3: Operation
BioTek Instruments, Inc.
If
SMP
is chosen:
- The transformation formula will be applied to the
sample
wells
only.
- SMP and any other well identifiers (
BLK, PC, NC, STD,
etc.)
defined will become available as MAP options when building the
transformation formula.
Example:
The assay plate map contains 2 NC wells and 2 PC wells. The
remainder of the map is filled with samples.
The assay data reduction requires that the mean of the NC be
subtracted from all the samples on the plate.
The transformation formula is:
SMP-NC;x
If
OD
is chosen:
- The formula definition screen will appear so that you can define a
formula for use
within
the transformation formula.
- Use the formula keys (Math, Other, Map and Function) to define
the Transformation Variable (TVAR). Once the variable has been
defined, it can be used in a transformation formula. The TVAR
will be available as a MAP option when building the
transformation formula.
Example:
The assay plate map has 2 blanks, 1 control well in duplicate
(
CTL1
), 1 negative control well in triplicate
(NC)
, and 5 standards
in duplicate
(STD1-STD5)
.
The assay data reduction states:
- Subtract the mean of
CTL1
from the mean of the
NC
. Subtract the
difference from all ODs on the plate.
- Divide the result of the above by the mean of the
NC
less the mean
of
CTL1
, and then multiply by 100.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...