Chapter 6: Troubleshooting and Error Codes
BioTek Instruments, Inc.
General Errors (Continued)
1100
Failed configuration checksum test
The stored checksum value from the configuration data does not match the
actual checksum of the current configuration data.
Probable Cause:
•
The configuration data has changed and the checksum stored is no longer
valid. The error is produced when outdated versions of assay definition
software are used to create an assay configuration file.
•
This file is incompatible with the operation code within the instrument’s
memory.
Recreate the assay definition using the correct version of assay
definition software and re-download the software.
1200
Autocalibration data missing
No Autocal data exists for the read location.
Probable Cause:
•
Autocalibration was not performed following assay definition download.
1300
Motor not homed successfully
If the motor is not correctly homed, the instrument sends an error and exits the
function.
Probable Causes:
•
Errors 0200 or 0300 have been ignored. Refer to
Probable Causes
for
these errors on pages 140 and 141.
Fatal Errors
Fatal Errors indicate conditions that require immediate attention. If a fatal error is
displayed, contact BioTek’s Technical Assistance Center for further instructions.
Refer to
Chapter 1
for contact information.
Code Description and Probable Causes
A100 Task control block not available
A200 Read already in progress
A300 Device not available
A304 Printer device not available
A400 Failed code checksum on power-up
A500 Code flash readback timed out
A600 Quick flash configuration timed out
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...