Qualification Procedures
|
111
ELx800 Operator's Manual
Defining the Absorbance Test Plate Parameters
Using the Standards Certificate provided with the Absorbance Test Plate, follow
the steps below to define the Test Plate parameters.
To define the Test Plate parameters via the instrument keypad:
1.
Start at the Main Menu and select
UTIL SELECT UTILITY
OPTION SETUP
. The
EDIT SETUP INFORMATION
menu will appear:
E D I T S E T U P I N F O R M A T I O N :
D A T E T I M E F I L T E R * M O R E
2.
Press
*MORE
to display the second
EDIT SETUP
menu. Select
CALPLATE
to
access the
CALIBRATION FILTER
menu.
3.
Using the Standards Certificate packaged with the Test Plate, select the
filter wavelength.
E D I T S E T U P I N F O R M A T I O N :
R S 2 3 2 C A L P L A T E
* M O R E
C A L I B R A T I O N F I L T E R : 4 0 5
4 0 5 4 5 0 4 9 0 * M O R E
4.
Press
ENTER.
The
WELL LOCATION/CALIBRATION VALUE
screen
appears:
W A V E L E N G T H : 4 0 5 W E L L : C 0 1
C A L I B R A T I O N V A L U E : 0 . 1 5 8
5.
Enter the values listed on the Test Plate Standards Certificate. After each
entry, press
ENTER
to advance to the next consecutive well location.
Repeat for the remaining filters.
6.
When all values have been entered, press the Main Menu key.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...