![Biotek EL800 Operator'S Manual Download Page 138](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074138.webp)
112
|
Chapter 4: Instrument Qualification
BioTek Instruments, Inc.
To define the Test Plate parameters in Gen5™:
Gen5’s
Reader
Diagnostics Utility
(PN 5320201) needs to be
installed
before
you can perform the Absorbance Plate Test in Gen5.
1.
Select
System
|
Diagnostics
|
Test Plates|Add/Modify Plates
, and click
Add
. Click the
Help
button for guidance.
To define the Test Plate parameters in KC4™:
1.
Select
System
|
Diagnostics
|
Define Universal Plates|Add
. Click the
Help
button for guidance.
To define the Test Plate parameters into KCjunior™:
1.
Select
Utilities
|
Diagnostics
|
Universal Plate Test
.
2.
Click
New Data Sheet
. Click the
Help
button for guidance.
Running the Absorbance Plate Test
IMPORTANT:
Before running the Absorbance Plate
Test, ensure that the reader is not running in
Rapid
Mode. To check this, select
UTIL READ
and then
cycle through the prompts until
READ IN RAPID
MODE?
is displayed. Choose
NO
for an accurate result.
To run the Absorbance Plate Test from the instrument keypad:
1
Start at the Main Menu and select
UTIL TESTS CALPLATE
. The
SELECT TEST
menu will appear:
S E L E C T T E S T :
S Y S T E M C H K S U M
C A L P L A T E
2
Select the appropriate wavelength at the
CALIBRATION FILTER
screen,
and press
ENTER
to save the value and continue.
C A L I B R A T I O N F I L T E R : 4 0 5
4 0 5 4 5 0 4 9 0 * M O R E
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...