Chapter 3: Operation
BioTek Instruments, Inc.
C U R V E - F I T T Y P E : C - S P L I N E
C - S P L I N E
P T - P T * M O R E
•
Linear curve fit:
A simple best-fit straight line is plotted using the
values of the standards.
•
Quadratic or “Quad” curve fit:
A curve fit that uses the quadratic
equation “ax
2
+bx +c = y” to plot the standard’s values. Utilizing this
curve, any data point for a standard that deviates from the ideal value
will not affect the entire curve.
•
Cubic curve fit:
A curve fit that uses the equation
“ax
3
+ bx
2
+ cx + d = y” to plot the standard’s values. This type of curve
fit is affected even less than the quadratic fit when any particular
standard has a poor value.
•
2-P (LOGIT/LOG):
A curve fitted to the standard values, which is
characterized by a skewed sigmoidal (S-shaped) plot that eventually
becomes asymptotic to the upper and lower standard values. The
logistic equation is algebraically transformed to a simpler form in which
experimentally determined values are used for the responses at
concentrations of zero and infinity.
Note:
Uses linear axis type only.
•
Cubic Spline (C-Spline) curve fit:
A piecewise polynomial
approximation consisting of joining a set of data points by a series of
straight lines, which is then smoothed by using a cubic fit.
•
4-Parameter Logistic or “4-P”:
A curve fitted to the standard values,
which is characterized by a skewed sigmoidal (S-shaped) plot that
becomes asymptotic to the upper and lower standard values. The 4
parameters are: Left asymptote, Right asymptote, Slope and Value at
the Inflection point. This fit is most recommended for immunoassay
data, and is more exact than Logit/Log.
Note:
Uses linear axis type
only.
•
Point to Point or “PT to PT”:
A plot that connects each standard point
with a line, with no averaging of the values to “smooth” the curve at
each standard.
•
Press
SOFT KEYS 1, 2, 3,
or
4
to select the curve-fit type that is
displayed above the soft key. Select
*MORE
to display additional
options.
•
Press
ENTER
to save the selection and advance to the next screen.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...