Results Interpretation
The association between low positive levels of anticardiolipin antibodies and
clinical findings is unclear.
Normal population studies indicate that there is a higher prevalence of IgM
positives in the normal population than IgG, 9.4% and 6.5%, respectively. In
normal pregnancy, the levels are higher still at 17.0% (IgM) and 10.6% (IgG).
Expected Values
The normal range was determined on serum from 102 normal adult blood donors.
The ranges below are provided as a guide only. ELISA assays are very sensitive
and capable of detecting small differences in sample populations. It is
recommended that each laboratory determine its own normal range, based on the
population techniques and equipment employed.
IgG Anticardiolipin
< 11 GPL U/ml
Negative result
> 11 GPL U/ml
Positive result
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...