18
|
Chapter 2: Installation
BioTek Instruments, Inc.
Power-Up and System Test
After you have installed the ELx800™ and connected the power supply, turn on
the instrument to run a system test. The on/off switch is located on the lower right
side of the base.
The
System Test
begins with a check of the stepper motors and the analog power
supplies, to ensure that they have a proper input voltage level. The data flash
checksum, motor axis, and analog offset are then verified. The photodetector’s
dark current, noise, and gain are checked to ensure they fall within specific
pass/fail criteria.
If an error is detected, the reader will “chirp” and display an error code. See
Chapter 6
for a list of error codes. If no errors are detected, the reader will briefly
display
SYSTEM TEST PASS
.
The power-up system test does not produce a printed results report. To run the
test manually and obtain a printout of the system test values, start at the Main
Menu and press
UTIL
TESTS
SYSTEM
.
See
System Test
and
Checksum
Test
in
Chapter 4
, or
Connecting a Printer to the ELx800
on page 20 for more
information.
ELx800 Main Menu
Following successful power-up of the ELx800, the
Main Menu
appears:
R E A D Y
0 1 : 2 2 P M
0 1 / 2 3 / 0 4
R E A D
D E F I N E R E P O R T U T I L
•
The
Main Menu
permits access to all onboard functions. See
Main Menu
in
Chapter 3
for more information, including a diagram showing the flow of
functionality (
Figure 15
).
•
The ELx800 front panel contains four circular buttons, referred to in this
manual as
“SOFT KEYS.”
One
SOFT KEY
is positioned directly below
each selectable option in the display. To select a menu option, simply press
its corresponding
SOFT KEY
. See
ELx800 Front Panel
in
Chapter 3
for
additional instructions.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...