S12 Vet Monitor User’s Manual
52
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
ECG Cable will
expire soon
Prompt
/
ECG cable is about to expire. Please
replace the cable in time.
ECG Suspected
Pacing Signal
Prompt
/
Pacing signal has been detected by
non-pacing patients. Please check
whether the patients have pacemakers
and check the connection of ECG
electrode sheets.
Note: XX represents RA, LA, LL, RL, V1, V2, V3, V4, V5, V6,
YY represents I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5 or V6.
D.2.3 RESP Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
RESP Leads Off
Med
Completely
clear
ECG lead-off or the ECG cable is not
connected. Check the communication of
ECG electrode and lead wires.
D.2.4 SpO2 Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
SpO2 Self-test Error
Med
Not clearable
Board failure. Please contact the
manufacturer for repair.
SpO2 Sensor Off
Med
Sound and light
can be cleared
The SpO2 sensor is falled off from
the patient end. Check the
connection of the sensor. If the
alarm still exists, replace the
sensor.