S12 Vet Monitor User’s Manual
42
Item
Default Setting
Low limit
ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure: 35
mmHg
PA/PAWP: 12 mmHg
CVP/ICP/RAP/LAP/UVP venous pressure: 0 mmHg
Alarm priority
MED
Alarm output
OFF
CPP
Alarm switch
ON
High limit
90 mmHg
Low limit
30 mmHg
Alarm priority
MED
Alarm output
OFF
Unit
ART/Ao/UAP/BAP/FAP/LV/RAP/LAP/UVP/PA/PAW
P/P1/P2: mmHg
CVP /ICP/CPP: cmH2O
Sensitivity
MED
Waveform Speed
25 mm/s
Scale type
Manual
Scale
Upper scale
ART /Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure:
160mmHg
CVP /ICP/RAP/LAP/UVP venous pressure : 20mmHg;
PA/PAWP
:
30mmHg
Lower scale
0mmHg
High-precision
cursor
switch
OFF
High-precision cursor
ART/ Ao/ UAP/BAP/FAP/LV/P1/P2 arterial pressure:
80mmHg
CVP /ICP/RAP/LAP/UVP venous pressure: 10mmHg
PA/PAWP
:
15mmHg
PPV measurement
OFF
PPV source
Auto