S12 Vet Monitor User’s Manual
7-8
2.
Select
【
Alarm
】
→
【
Sound
】
.
3.
Set
【
Auto Increase Volume
】
to
【
3 Steps
】
,
【
2 Steps
】
,
【
1 Steps
】
or
【
Off
】
.
【
3 Steps
】
:
After the alarm occurs, the alarm volume will be
automatically increased to level 3, if the set time is not confirmed.
【
2 Steps
】
: After the alarm occurs, the alarm volume will be
automatically increased to level 2, if the set time is not confirmed.
【
1 Steps
】
: After the alarm occurs, the alarm volume will be
automatically increased to level 1, if the set time is not confirmed.
【
Off
】
: After the alarm occurs, the set time is not confirmed, and the
alarm volume remains unchanged.
4.
Set
【
Increase Volume Delay
】
, select sound enhanced delay time.
7.6.4.
Setting Alarm Delay Time
For the over-limit alarm of continuous measurement parameters, the alarm delay
time can be set. If the condition of triggering alarm disappears within the delay time, the
monitor will not alarm.
Set the delay time for the alarm by following these steps:
1.
Select
【
Main Menu
】
quick key→from
【
System
】
column to select
【
Maintenance
】
→input maintenance password→Enter.
2.
Select
【
Alarm
】
submenu→
【
Other
】
submenu.
3.
Set
【
Alarm Delay
】
.
The delay time of the apnea alarm is not affected by the alarm delay time setting.
You can set the delay time of the apnea alarm separately.
7.6.4.1.
Setting Apnea Alarm Delay Time
Steps to set apnea alarm
delay time are as follows:
1.
Enter
【
Setup
】
interface in either of the following ways:
Select
【
Alarm Setup
】
quick key→Select
【
Setup
】
submenu.
Select
【
Main Menu
】
quick key→from
【
Alarm
】
column to select