S12 Vet Monitor User’s Manual
15-10
——
【
O
2
Compensation
】
: Select the concentration of oxygen. It can be set to a
value between 0% and 100%. The default value is 16%.
——
【
Anesthetic Gas
】
: Select the concentration of anesthetic agent. It can be set to
a value between 0.0% and 20.0%. The default value is 0.0%.
——
【
Balance Gas
】
:
Select the type of balance gas, the options are
【
Room
Air
】
,
【
N
2
O
】
or
【
Helium
】
.
When the most proportions of the mixture is air, select
【
Room
Air
】
When the most
proportions of the mixture is N
2
O, select
【
N
2
O
】
.
When the most proportions of the
mixture is Helium, select
【
Helium
】
.
The Masimo mainstream/sidestream CO
2
module supports automatic air pressure
compensation and automatic temperature compensation. Manual setting is not required
under normal conditions, but other gas interferences may exist in the gas circuit. In order
to ensure accurate measurement of Masimo module, the following compensations can be
manually set according to actual conditions:
——
【
O
2
Concentration
】
: Set the oxygen concentration. Can choose 0% ~ 100%,
the default value is 16%.
——
【
N
2
O Concentration
】
: Set the concentration of N
2
O. Can choose 0% ~ 100%,
the default value is 0%.
WARNING:
Please set the CO
2
corrections according to actual situation, otherwise, the
measured value may be inaccurate and away from actual value.
15.6.6.
Operating mode
You can select the operating mode of the CO
2
module according to the actual
situation of the module. The operation steps are as follows:
1.
Select the CO
2
parameter area or waveform area to enter the
【
CO
2
】
menu.
2.
Select
【
Setup
】
submenu.
3.
Set
【
Operation Mode
】
.
Measure: Select measurement mode is required when measuring with the CO2
module.