S12 Vet Monitor User’s Manual
23-1
Chapter 23
Maintenance
WARNING:
Failure on the part of the responsible individual hospital or institution
employing the use of this equipment to implement a satisfactory maintenance
schedule may cause undue equipment failure and possible health hazards.
The safety checks or maintenance involving any disassembly of the equipment
should be performed by professional servicing personnel. Otherwise, undue
equipment failure and possible health hazards could result.
If necessary, please contact the manufacturer for product circuit diagrams,
parts lists, calibration instructions or other equipment maintenance related
information.
If there is a problem with the monitor, please contact the maintenance
personnel or us.
23.1.
Inspection
Before use, after continuous use for 6-12 months, maintenance or upgrade, qualified
maintenance personnel should conduct a comprehensive inspection to ensure the normal
operation and work of the monitor.
Items to be inspected shall include:
The environment and power supply meet the requirements.
There is no mechanical damage to the monitor and accessories.
The power cord has no abrasion and good insulation performance.
Use the specified accessories.
The alarm system functions normally.
The recorder works normally and the recording paper meets the specified
requirements.
The performance of the battery.
Various monitoring functions are in good working condition.
Grounding impedance and leakage current meet the requirements.