S12 Vet Monitor User’s Manual
8-10
1.
Select the ECG
parameter area or waveform area to enter the
【
ECG
】
menu.
2.
Select
【
Alarm
】
submenu.
3.
If the alarm settings are password protected, enter the password. For details,
please refer to
7.6.2 Changing Alarm Setup Protection Mode.
4.
Set alarms as needed.
8.5.3.
Setting ECG calculating Lead
You can set the label name of the ECG calculation lead as follows:
1.
Select ECG parameter area or waveform area to enter the
【
ECG
】
menu.
2.
Select
【
Setup
】
submenu.
3.
Select
【
ECG 1
】
or
【
ECG 2
】
to set label name of ECG calculating Lead.
WARNING
:
Only when you switch on
【
MultiLead Ana
】
can you set
【
ECG 2
】
.
8.5.4.
Setting Multi-lead Analysis
When multi-lead analysis function is switched on, the
【
ECG 2
】
participate in the
calculation of HR, the steps to set up the multi-lead analysis switch are as follows:
1.
Select ECG
parameter area or waveform area to enter the
【
ECG
】
menu.
2.
Select
【
Setup
】
submenu.
3.
Switch on or off
【
MultiLead Ana
】
.
CAUTION
:
【
ECG 1
】
is the key calculation lead;
【
ECG 2
】
is the auxiliary calculation lead.
Only when the ECG
【
Lead Type
】
is 5/6/12 lead can you set
【
MultiLead Ana
】
.