S12 Vet Monitor User’s Manual
30
IEC 61000-3-3
If the system is operated within the electromagnetic environment listed in Table
Guidance and Declaration – Electromagnetic Immunity, the system will remain safe and
provide the following essential performance:
Operating mode
Accuracy
Function
Accessories identification
Date stored
Alarm
Detect for connection
Parameter
Table 2
Guidance and manufacture’s declaration – electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of monitor should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment
– guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic
material,
the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines
2kV for power
supply lines
1 kV for
input/output lines
Mains power quality should
be
that
of
a
typical
commercial
or
hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth