S12 Vet Monitor User’s Manual
1-2
1.
Display screen
2.
Light induction: For example, when the environment is dark, the display brightness
can be automatically adjusted.
3.
Alarm lamp
Alarm lamp with different color and flashing frequency indicates the level of
technical alarm and physiological alarm:
High level alarm: the lamp quickly flashes red.
Medium level alarm: the lamp slowly flashes yellow.
Low level alarm: the lamp lights cyan without flashing.
4.
Power indicating lamp
It is a LED that lights green and orange, the status of the LED is specified as follows:
Green: When the AC mains is connected.
Orange: When the AC mains is not connected and monitor is powered by
battery.
Off: When the AC mains is not connected.
5.
Charging indicating lamp
Light up: When the battery is being charged.
Off: When the battery is fully charged or no battery in monitor.
6.
Phonetic holes.