S12 Vet Monitor User’s Manual
56
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
pressure is out of range or the
pressure is unstable.
XX Catheter Off
HIGH
Not clearable
The catheter is pulled out from the
patient. Please check the
connection.
Zero Required
Prompt
/
/
Zero Succeed
Prompt
/
The IBP module is zeroing
successful
Note: XX represents IBP labels, such as PA, CVP, FAP, P1, etc.
D.2.8 C.O. Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
C.O. TB Self-test
Error
Med
Not clearable
The board is failure. Please
contact the factory for repair.
C.O. TI Self-test
Error
Med
Not clearable
The board is failure. Please
contact the factory for repair.
C.O. TB Sensor Off
Med
Completely clear
Check the sensor connection and
reconnect the sensor.
C.O. T1 Sensor Off
Med
Completely clear
Check the sensor connection and
reconnect the sensor.
C.O. Measure
Timeout
Low
Completely clear
When measuring manually, it do
not inject for sustaining 20
seconds.
D.2.9 CO2 Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution