S12 Vet Monitor User’s Manual
26
Alarm limit
Range
Step
NIBP-S-High
(low limit+1 mmHg) ~270 mmHg
1mmHg
NIBP-S-Low
30 mmHg ~ (high limit-1 mmHg)
NIBP-M-High
(low limit+1mmHg) ~235 mmHg
1mmHg
NIBP-M-Low
20 mmHg ~ (high limit-1 mmHg)
NIBP-D-High
(low limit+1mmHg) ~220 mmHg
1mmHg
NIBP-D-Low
10 mmHg ~ (high limit-1 mmHg)
A.9.4 SpO2
Alarm limit
Range
Step
SpO2 High
(low limit+1
%
) ~100
%
1
%
SpO2 Low
(SpO2 Desat +1
%
) ~ (high limit-1
%
)
SpO2 Desat
0
%
~ (low limit-1
%
)
PR High
(PR low limit+1bpm)~400bpm
1bpm
PR Low
10bpm~ (PR high limit-1bpm)
A9.5 TEMP
Alarm limit
Range
Step
T1/T2 High
(low limit+0.1ºC) ~50.0ºC
0.1 ºC
T1/T2 Low
0 ºC~ (high limit-0.1ºC)
0.1 ºC
TD High
0ºC~5.0ºC
0.1 ºC
A.9.6 IBP
Alarm limit
Range
Step
IBP-M-High
(low limit +1mmHg) ~360mmHg
1mmHg
IBP-M-Low
-50mmHg~ (high limit -1mmHg)
IBP-D-High
(low limit +1mmHg) ~360mmHg
1mmHg
IBP-D-Low
-50mmHg~ (high limit -1mmHg)
IBP-S-High
(low limit +1mmHg) ~360mmHg
1mmHg