S12 Vet Monitor User’s Manual
23
monoxide)
5)
NO (Nitrogen
monoxide)
5)
0.02 vol%
–
1)
–
1)
–
1)
–
1)
O2
5)
100 vol%
–
2)
–
2)
–
1)
Note 1:
Negligible interference, effect included in the specification “Accuracy, all
conditions” above.
Note 2:
Negligible interference with N2O / O2 concentrations correctly set, effect
included in the specification “Accuracy, all conditions” above.
Note 3:
Interference at indicated gas level. For example, 50 vol% Helium typically
decreases the CO2 readings by 6%. This means that if measuring on a mixture containing
5.0 vol% CO2 and 50 vol% Helium, the actual measured CO2 concentration will typically
be (1-0.06) * 5.0 vol% = 4.7 vol% CO2.
Note 4:
According to the EN ISO 80601-2-55:2011 standard.
Note 5:
In addition to the EN ISO 80601-2-55:2011 standard.
Respironics CAPNOSTA 5
Gas or Vapor
Gas Level
Nitrous oxide
60%
Halothane
4%
Enflurane
5%
Isoflurane
5%
Sevoflurane
5%
Xenon
80%
Helium
50%
Metered dose inhaler propellants
Unspecified
Desflurane
15%
Ethanol
Unspecified
Isopropanol
Unspecified
Acetone
Unspecified
Methane
Unspecified
Additional notes regarding cross-sensitivity compensation errors:
Xenon: The presence of Xenon in the exhaled breath will negatively bias Carbon Dioxide