Battery
21.1.
Introduction
The monitor can be fitted with rechargeable battery to ensure the normal use of
the monitor in case of intra-hospital patient transfer or whenever the power supply is
interrupted. When the monitor is switched on with AC power, the battery can be
charged regardless of whether the monitor is switched on or not. Since we do not
provide external charging equipment, the battery can only be charged in the monitor.
In case of sudden power failure, the system will automatically use battery to supply
power to the monitor, thus not causing interruption of monitoring work.
On-screen battery symbols indicate the battery status as follows:
Indicate that the battery works correctly. The solid portion represents the
battery’s charge.
Indicate that the battery is almost depleted and needs to be charged
immediately. Otherwise, the monitor will be automatically shut down. In
this case, the monitor sends out an alarm message.
Indicate that no battery is installed.
The power supply of battery can only function for a certain period. Excessively
low voltage of battery will trigger a high priority technical alarm
【
Battery Low
】
. At
this moment, the monitor shall immediately connect with alternating current power
supply to charge the battery.
In case of long-term monitoring, a backup battery shall be installed and used after
the AC power is plugged in. The AC power plug must be plugged into the special
interface of the hospital.
21.2.
Installing a Battery
The battery of this monitor must be installed and replaced by maintenance personnel
trained and authorized by our company.
21.3.
Battery Guidelines
The service life of the battery depends on the frequency and time of use. If