S12 Vet Monitor User’s Manual
50
Alarm messages
Default priority Cause and solution
probe.
SpO2 Desat
High
SpO2 measurement is below the desaturation
limit. Please check the patient’s status and
confirm whether the alarm limit setting is
applicable to the patient.
D.1.5 CO2 Physiological Alarm
Alarm messages
Default priority
Cause and solution
CO2 Apnea
High
The patient is not breathing or the
respiratory signal is too weak to measure
the respiratory rate. Check the patient’s
condition and whether the air circuit
connection is correct.
D.2 Technical Alarm
This chapter lists the main technical alarms, the level of technical alarms, the cleared
status of alarm reset alarm prompts, and the measures to be taken after the alarm occurs.
Some alarm messages may not be listed.
After different technical alarm is reset, the alarm prompt will be cleared to different
degrees. The following three types of technical alarms are given in this section according
to the status of alarm being cleared.
Completely clear: the technical alarm is completely clear. The monitor has no
alarm indication.
Sound and light can be cleared: the technical alarm displays as prompt
information.
Not clearable: the sound of technical alarm is shielded.