S12 Vet Monitor User’s Manual
15
Sensor calibration
time
One year (recommend)
Unit
mmHg, kPa
Dynamic
pressure
measurement range
Systolic
30~270 mmHg
Diastolic
10~220 mmHg
Mean
20~235 mmHg
Dynamic
Pressure
Measurement Error
of Simulator
Systolic
35~255mmHg: ±8 mmHg
Diastolic
15~195mmHg: ±8 mmHg
Mean
22~215mmHg: ±8 mmHg
Static pressure
accuracy
±3 mmHg
(
±0.4kPa
)
Pressure resolution
1 mmHg or 0.1kPa
PR
measurement
range and accuracy
Measurement
range
40 ~ 240 bpm
Accuracy
±3bpm or ±3%, whichever is greater
Maximum
measurement time
<120s
First overpressure
protection point
297±3 mmHg
Second overpressure
protection point
315±10 mmHg
Note:
The accuracy of NIBP cannot be determined by using a simulator, but under many
conditions, it is still necessary to use a simulator to test its performance (for example, a
simulator is required for quality control in the production process), and the simulator of
the model specified by the manufacturer shall be used for this performance test.
A.8.4 SpO2
A.8.4.1 Standard
Refer to the standard of ISO 80601-2-61