S12 Vet Monitor User’s Manual
11
Calibration voltage
≤±5%
(
×1
)
Input offset current
<
0.1uA
Time constant
Monitoring mode: ≥0.3s
Diagnostic mode: ≥3.2s
A.8.1.4 ECG analysis calculation
Measurement
parameter
HR, PVCs, ST, QT, and Arrhythmia analysis
Multi-lead
synchronous
analysis function
Supports synchronous analysis of at least 2 leads, one of which
is the key monitoring lead and the other is the auxiliary lead. It
is on except 3-lead mode.
Smart Lead Switch
Automatic, Manual; Default manual
(3-lead mode is fixed as manual)
Automatic mode: the algorithm automatically identifies the
current smart leads, and the host automatically switches the key
monitoring leads according to the identification of the algorithm.
HR
measurement
range and accuracy
Measurement
range
10~400 bpm
Resolution
1 bpm
Accuracy
±1% or ±1 bpm, whichever is greater
Detection
sensitivity
0.20mVp-p
ST Display
Display 12-lead ST segment values at the same time and support
ST graphic display.
ST
measurement
range and accuracy
Measurement
range
-2.0mV~2.0mV
Resolution
0.01mV
Accuracy
-0.8mV~0.8mV:
±0.02mV
or
±10%,
whichever is greater
Other: Unspecified
ST Update period
10s