S12 Vet Monitor User’s Manual
54
D.2.6 NIBP Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
NIBP Self-test
Error
Med
Not clearable
Board failure. Please contact the
manufacturer for repair.
NIBP System
Failure
Low
Not clearable
System operation failure.
NIBP Air Pressure
Error
Low
Completely
clear
Pressure error, unable to maintain stable
cuff pressure, such as tracheal knot.
NIBP
Air Leakage
Low
Completely
clear
NIBP air leakage was found in the
inspection. Please check the sleeve and
airpipe for air leakage.
NIBP Air System
Leak
Low
Completely
clear
Damaged cuff, hose or joint.
NIBP Cuff Type
Error
Low
Completely
clear
The cuff used does not match the patient
type set. Please confirm that the patient
type is set correctly and select correct
cuff according to the patient type. If the
patient type and cuff selection are
correct, please check whether the airway
and airpipe are bent or blocked.
NIBP
Overpressure
Detected
Med
Completely
clear
The pressure exceeds the specified
safety limit.
NIBP Loose Cuff
Low
Completely
clear
The cuff is not tight; Or the cuff is not
connected. Select the correct cuff
according to the patient type, place the
cuff according to the manual, and
connect the airpipe.
NIBP Excessive
Motion
Low
Completely
clear
The patient moved frequently during the
measurement. Or violent movement
during measurement;