S12 Vet Monitor User’s Manual
19-4
19.6.4.
Setting the duration of period recording
You can set the duration of every period recording in the following ways:
1.
Open the
【
Record Setup
】
menu.
2.
Set
【
Cycle Record Duration
】
to:
【
8s
】
: Record the waveform of 4 seconds before and after the current
time.
19.6.5.
Setting the recording speed
1.
Open the
【
Record Setup
】
menu.
2.
Set
【
Record Speed
】
.
This setting is applicable to all recording tasks with waveforms.
19.6.6.
Setting alarm recording duration
You can set how long the waveform needs to be recorded when an alarm occurs, as
follows:
1.
Open the
【
Record Setup
】
menu.
2.
Set
【
Alarm Record Duration
】
.
【
8s
】
: Record the waveform of 4 seconds before and
after the alarm triggering time.
19.6.7.
Setting NIBP Trigger Record
You can set to record the output NIBP measurement results when NIBP
measurement is completed, as follows:
1.
Open the
【
Record Setup
】
menu.
2.
Set
【
NIBP Trigger
】
to
ON
or
OFF
.
19.7.
Installing Recording Paper
If the record paper runs out, please install the record paper as the following step:
1.
Press both sides of the recorder door with one hand and pull outwards to open