S12 Vet Monitor User’s Manual
16
A.8.4.2 Specification
Measurement parameter
SpO2, PR, PI and RR, SpO2 waveform and respiration
waveform.
Sensitivity
High, Medium, Low
Measurement range and
accuracy
Measurement range
0~100%
Simulator measurement
error
70%~100%: ±2%
0~69% unspecified
SpO2 Update period
Normal
≤2s
Maximum
≤25s
SpO2 Response time
Sensitivity
High
≤8s
medium
≤11s
Low
≤15s
PR
Measurement range
25 bpm ~
4
00 bpm
Resolution
1 bpm
Accuracy
± 3bpm
PI
Measurement range
At least 0.05~20.00%
Resolution
0.01%
Accuracy
±0.1% or ±10% of reading,
whichever is greater
RESP (from pleth)
Measurement range
0 rpm ~90 rpm
Resolution
1 rpm
Accuracy
±2 rpm
A.8.5 TEMP
A.8.5.1 Standard
Refer to the standard of ISO 80601-2-56.