S12 Vet Monitor User’s Manual
8-21
the same as that of ECG waveform, usually green. ST baseline segment and baseline
value are white.
You can select the ST waveform area to enter the
【
ST View
】
page or enter the
【
ST
View
】
page through the following steps:
1.
Select ECG parameter area, waveform area or ST parameter area to enter
【
ECG
】
menu.
2.
Select
【
ST
】
submenu.
3.
Select
【
ST View
】
from the bottom of the menu.
8.7.6.
Saving the ST Baseline
ST analysis requires valid samples. Set an ST baseline when ST values become stable.
If you do not set a baseline, the monitor will automatically save a set of baselines about 5
minutes after a valid ST measurement appears. You can also manually update the baseline
by selecting
【
Set Baseline
】
in the lower left corner of the
【
ST View
】
interface.
You can also make the following settings under the ST interface:
Select
【
Show Baseline
】
or
【
Hide Baseline
】
to show or hide ST baseline
segments and parameter values.
Select
【
Show Mark
】
or
【
Hide Mark
】
to show or hide ST reference point,
J point and ST point positions.
CAUTION
:
Changing the ST baseline will affect the ST alarm.
8.7.7.
Entering ST Graphic Window
The steps to enter the ST Graphic window are as follows:
1.
Select ECG parameter area, waveform area, ST parameter area or ST
waveform area to enter
【
ECG
】
menu.
2.
Select
【
ST
】
submenu.
3.
Select
【
ST Graphic
】
from the bottom of the menu.