S12 Vet Monitor User’s Manual
14
RR
measurement
range and accuracy
Measurement range
0~150 rpm
Resolution
1 rpm
Accuracy
±2 rpm or ±2%, whichever is greater.
Bandwidth
0.2 Hz ~2.5Hz
Sweep speed
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s, ±10%
Baseline impedance
range
200~2500Ω (using defibrillator proof cable with resistance of
1kΩ)
Measuring impedance
range
0.3Ω~3Ω
Gain
×0.25, ×0.5, ×1, ×2, ×4
A.8.3 NIBP
A.8.3.1 Standard
Refer to standard of IEC80601-2-30
A.8.3.2 Measurement Specification
Measurement
parameters
SYS, DIA, MAP,PR
Mode of operation
Manual, Auto, STAT, Sequence
Intervals for periodic
measurement time
1min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min,
30 min, 1h, 1.5h, 2h, 3h, 4h, 8h
Sequence mode
At least 5 groups are supported, and each group individually
sets the interval and number of periodic measurement.
STAT mode cycle
time
5 min
Measurement range
of cuff pressure
0~300 mmHg
Initial
inflation
pressure
Big animal: 120~280mmHg
,
default 160mmHg
Small animal: 60~280mmHg
,
default 160mmHg