NEW STYLE HGMSZ TECHNICAL SUPPORT NOTES 
FOR ALL SERIAL NUMBER STARTING WITH NU XXX AND ABOVE (JULY 2008) 

 

<4000> Zero Fault – IR Voltage out of tolerance 
 
<4200> Zero Fault and Gain set Fault 

 

<8000> Clipping Fault – Detector voltage out of tolerance 

 
 
For both Zero Fault and Clipping Fault – Press the UP arrow key to locate the screen displaying 
the following- 
 
 

SILENCE ZONE 

TMP 

SQUELCH CAL 

 
 
 
 
Place curser on CAL – press right arrow key and ENTER at the same time. 
The new screen will show the revision number. Note the revision number. Press the 
SILENCE/QUIT key: the next screen will display the following –  
 
 

DIGIPOT MODEL 
P CAL  

IR 

 
 
 
Place the curser on DIGIPOT – Press the ENTER key. The following screen will appear – 
NOTE – To verify correct model number, put curser on model and press enter. Standard 
refrigerant – 3015-4200, ammonia – 3015-4280. 
 
 

DIGIPOT = XXX 
IR VOLT = X.XXX 

 
 
 
Adjust the IR reading using the up and down arrow until the IR VOLT = 4.245 volts or as close 
as you can get it to that reading. Press SILENCE/QUIT key until you return to the screen reading 
PURGE. 
NOTE – The fault may not clear if the Digipot voltage is more then 0.15 volts away from 4.2 
volts. 

 
 
 
 

Revised 6/5/08 