PRESSURE READINGS FOR THE HGMSZ 
 
Codes 800, 1000 and 1800 

 
FIRST DIAGNOSTIC SCREEN 
 
 
 
                                                              Zone Pressure (0.06 psi below ambient) Maximum  

 

 

 

 

 

 

 2.7 PSIA   

4.26500V <0000> 
 
24.5Cd  

14.00psi 

Press left arrow key 
 
 

Revised 6/5/08 
P. M. Pape 

 
 
                                                                  Purge Pressure (0.03 psi below ambient) Maximum 

      

 

 

 

 

 

2.7 PSIA    

4.26500v <0000> 
 
24.5cD * 

14.3psi 

 
  Look for asterisk  
 
 
Press left arrow key until asterisk clears. Then press bottom arrow key. This will stop the pump. 
Take the pressure reading with the pump off then press the bottom arrow key to restart the pump. 
DO NOT LEAVE THE PUMP OFF FOR A LONG PERIOD OF TIME. 
 
 
 
 
                                                                   Ambient Pressure (pump off)     

4.26500v <0000> 
 
24.5cD  

14.50psi 

 
 
Locate the screen the shows as below. 
 
 

 

 

 

 

Place the curser on P-CK and press ENTER. 

P-CK  

CUST 

K2 

 
CUST K1 

CUST K3 

 
 
 
 
The next screen shows as below. 

Place your finger over the intake of the charcoal filter. The  

 

 

 

DIF reading should go to 2.00 or greater. 

13.11 * 

AMB 14.42 

 
1.21 DIF 

<0000> 

If this does not  occur then the pump is defective. 

 
 
Note: Asterisks (****) in V, MA and MW indicate an A to D failure on the main PC board. 
Note – Left arrow toggles valve. * equals purge pressure. No * equals zone pressure. 
 
 
 