The sensor triggers from high to low, and Remote Teach behaves electrically like trigger.
NOTE: If the device used to trigger or remote teach the iVu Plus BCR is a sinking device (that is, NPN), then
there are two options regarding the use of a pull-up resistor:
Option 1: Put a pull-up resistor, rated approximately 1k ohm, between the sensor's positive (+) voltage and
the sensor's input as shown below.
Trigger Input
Remote Teach Input
iVu
Pullup Resistor
1k ohm
NPN
VCC
Sinking
Option 2: Enable the Input Pullup in the iVu Plus BCR software ( Main Menu > System > Discrete I/O >
Input Pullup ).
iVu Output Waveforms
PNP Outputs
Ready
(Pass)
(Fail)
(Pass)
(Pass)
(Fail)
(Pass)
Function
Strobe Out
(Configure as External
& Expose Based)
Outputs 1,2, and 3
(Set as Active Pass)
Outputs 1, 2, and 3
(Set as Active Fail)
NPN Outputs
iVu Plus BCR User's Manual
Rev. B
www.bannerengineering.com - tel: 763-544-3164
13
Summary of Contents for iVu Plus BCR
Page 1: ...iVu Plus BCR User s Manual Rev B 1 25 2012...
Page 56: ...iVu Plus BCR User s Manual 56 www bannerengineering com tel 763 544 3164 Rev B...
Page 64: ...iVu Plus BCR User s Manual 64 www bannerengineering com tel 763 544 3164 Rev B...
Page 104: ...iVu Plus BCR User s Manual 104 www bannerengineering com tel 763 544 3164 Rev B...