Sterilization
17
EN
7
Sterilization
NOTE!
The sterilization of the video lens always follows the successful cleaning and dis-
infection.
WARNING!
Sterilization without prior cleaning and disinfecting is not allowed and may en-
danger the user, the patient, as well as third parties.
NOTE!
Do not use hot air sterilization.
7.1
Packaging
You can use standardized packaging material. The bag must be large enough for
the lens to prevent the seal from being stressed.
NOTE!
Make sure the product is sealed without wrinkles.
7.2
Moist heat sterilization
NOTE!
Perform the sterilization with moist heat (EN ISO 17556) in an autoclave with
fractionated pre-vacuum according to national requirements.
Steam sterilization
• Temperature retention time:
4 minutes at 134 °C (273 °F)
4 minutes at 132 °C (270 °F) (USA only)
18 minutes at 134 °C (273 °F)
• Drying time: 10-20 minutes
• Maximum sterilization temperature: 138 °C (280 °F)
WARNING!
Danger of burns! Video lenses are hot after steam sterilization. Allow sufficient
time for cooling to avoid burns.
NOTE!
When sterilizing several instruments with one sterilization cycle, make sure not
to exceed the max. load of the sterilizer.
Summary of Contents for AR-3210-0006
Page 8: ......
Page 10: ......
Page 30: ......
Page 32: ......
Page 54: ......
Page 76: ......
Page 96: ......
Page 98: ......
Page 118: ......
Page 120: ......
Page 140: ......
Page 142: ......
Page 162: ......
Page 164: ......
Page 184: ......
Page 186: ......
Page 206: ......
Page 208: ......
Page 228: ......
Page 230: ......
Page 252: ......
Page 274: ......
Page 294: ......
Page 296: ......
Page 316: ......
Page 318: ......
Page 338: ......
Page 340: ......
Page 360: ......
Page 362: ......
Page 383: ......
Page 384: ......