12
Visual Inspection and Function Testing
EN
Replace the video lens, if:
• the connection to the endoscope or the camera is not secure
• pressing zoom and focus adjustment ring does not result in image changes
• the picture quality is not sufficient despite cleaning the lens protection win-
dow
• sharp edges or deformation are seen on the video lens