Specifications
Sample
Measurement item
Measurement wavelength
Measurement principle
Measurement Range
Reagent Strip
Processing speed
Minimum sample volume
Sample consumption
Sample container
Simultaneous measurement
Reagent reaction temperature
Light source
Calibration method
Data storage volume
Display
Built-in Printer
External Output
Transmission Method
Transmission Rate
Measurement conditions
Power supply
Power input
Dimensions
Weight
Serum, Plasma, Whole blood (Only Hb)
General biochemical measurement items, 22 items
5 wavelengths (405, 550, 575, 610 and 820 nm)
Optical measurement of reflection intensity of
reagent color reaction
Set for sach measurement item
SPOTCHEM
Ⅱ
Reagent Strip
63 samples per hour
6
×
the number of measurement items + 38
μ
L:
serum, plasma
4
〜
6
μ
L(per 1 measurement item): serum, plasma
Whole blood: exclusive whole blood sample tube
(Orange cap)
Centrifuge tube
Serum, plasma: exclusive serum sample tube
(Blue cap)
3 Single Reagent Strips or 1 Multi Reagent Strip
Maximum of 9 test items of continuous
measurement using Single Reagent Strips and a
Multi Reagent Strip is available.
37
℃
LED and Interference filter
Calibration by magnetic card (Reagent Card)
Calibration by Calibrator Kit
100 tests
20 digits
×
2 lines LCD
36-character thermal printer (58 mm width)
RS-232C interface
Single or Two-way Transmission
300, 600, 1200, 2400, 4800, 9600bps
Temperature: 10
〜
30
℃
Humidity: 20
〜
80 % (Non-condensing)
AC100
〜
120V, 220
〜
240V
50/60Hz
(
AC adapter method
)
Max. 100VA
338 mm(W)
×
203 mm(D)
×
167 mm(H)
Approximately 5.4 kg
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...