b
SP-4430 ● 2-11
To obtain the appropriate
contrast on the screen, rotate
the contrast adjustment knob
(“CONT.”) on the rear of the
analyzer.
▲
1-4
装置の設置
Chapter 2 MEASUREMENT
2-3-2
Startup
The operating procedures from turning ON the power to the MAIN MENU
are described below.
Turn ON the power.
1.
Turn ON the power at the rear of the
analyzer. When the power is ON, “|”
is displayed
The name of the analyzer and the
system version (“VXXXX” shown at
the right) are displayed and warm-up
starts.
The screen appears during warm-up
as shown at the right, and initialization
and self-check of each function are
performed.
About 10 minutes after (when at
25
℃
), warm-up is completed and the
MAIN MENU is displayed.
●
●
●
●
Warm-up time depends on the
room temperature.
▲
S P O T C H E M E Z
S Pー4430 V X X X X
NOTE
Before turning ON the power, make sure that the maintenance cover is in
position. If light enters into the analyzer during initialization, an error will
occur.
Before turning ON the power, make sure that nothing is placed in front of
the Table Cover. The Table Cover opens during initialization. If there is an
obstruction, a problem may occur.
W a r m i n g u p・・・
/
1 . M e a s u r e 2 . S u b m e n u
3 . C a l i b r a t e
( 1/1 )
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...