Chapter 4 MAINTENANCE
4-1-1
お手入れの目安
Disinfect and wash the Tip Waste Case.
3.
Disinfect the Tip Waste Case with
alcohol and wash it with tap water.
Wipe it dry with a cloth.
●
●
Reattach the Tip Waste Case.
4.
Put the Tip Waste Case into the
Centrifuge-equipped Multi Rack.
●
End cleaning.
6.
Turn ON the power.
Warm-up starts. The Reagent Table
and Centrifuge-equipped Multi Rack
slide back to the original position
and the Table Cover closes.
If the operation is completed, turn
OFF the power after the MAIN
MENU is displayed.
●
●
Disinfect and wash the Protective Cover.
5.
Disinfect the Protective Cover with
alcohol and wash it with tap water.
Wipe it dry with a cloth.
●
●
W a r m i n g u p・・・
/
1 . M e a s u r e 2 . S u b m e n u
3 . C a l i b r a t e
( 1/1 )