1-10 ● SP-4430
1-4 Setting up the Analyzer
Chapter 1 INTRODUCTION
1-4-1
Cautions
Before setting up the
analyzer, read the
following notes and
always take proper
safety precautions.
CAUTION
10cm or more
Allow a space of 10 cm or more between
the rear of the analyzer and the wall.
Failure to do so may cause overheating.
Excessive load on the cable connection
may cause fire or incorrect measurement
results.
●
AC100-120V
AC220-240V
50 / 60Hz
Operate the analyzer with power of the
correct voltage and frequency.
Otherwise fire or damage may result.
●
Do NOT unnecessarily disassemble or
modify the analyzer. Such actions may
invite danger of exposure to pathogenic
microbes, and cause fire or damage.
●
Place the analyzer on a stable and level
surface free of vibration. Failure to do so
may damage the analyzer, correct
measurement results may not be obtained,
and injury may result. Do NOT place the
analyzer where there is a risk of falling.
●
To avoid electric shock and/or fire, use
the attached power cord to connect with
a power outlet. For details, contact your
distributer.
●
Connect the power plug directly to an
outlet, not via an extension cord or
power tap. The power supply for the
analyzer is 100 VA. Before turning on the
power switch, make sure that the total
input of devices connected to a
receptacle of the same circuit doesn
’
t
exceed 1500 VA (100V, 15A).
●
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...