C
SP-4430 ● 2-15
Chapter 2 MEASUREMENT
Using a centrifuge
Prepare a centrifuge.
Uncap the tube and pipette 250
μL
of
heparinized whole blood or plain blood into
the centrifuge tube.
The centrifuge tube includes Heparin as an
anticoagulant.
Close the cap tightly after setting a sample,
and invert it.
Use the sample after removing the cap.
Wipe out the sample attached to the outside
of centrifuge tube before placing it to the
centrifuge, otherwise it may spattering the
sample.
Place samples in the centrifuge ( see“ 2-4-1
Normal measurement”).
●
●
●
●
●
●
Remove any air bubbles or skin on the surface of
the sample. They may cause incorrect measurement
results.
Wipe out the sample attached to the outside of
centrifuge tube before placing it to the centrifuge,
otherwise it may cause spattering the sample .
IMPORTANT
A Centrifuge tube has 1 line
marked on it for 250
μL
.
▲
The samples may coagulate when it takes time
before the samples collected are brought into the
machine. Those showing heavy coagulation or
deposition of fibrin sometimes can not be measure.
A syringe containing heparin can be used to
prevent coagulation when it takes time before the
samples are brought into the centrifuge, although
the machine contains heparin.
IMPORTANT
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...