C
5-06 ● SP-4430
4-2
毎日のお手入れ
Chapter 5 TROUBLESHOOTING
4-2-2
チップ廃却ケースの清掃
Description and error message
Conditions and causes
Remedy
The memory storing the measured
results is defective.
Power was cut while writing in memory
or deleting.
●
●
If the same trouble occurs often,
contact your distributor.
●
E90
E90 Memory:results
■■■■■■■■■■■■
OK(ENTER)
The memory storing the trouble history
is detective.
Power was cut while writing in memory
or deleting.
●
●
If the same trouble occurs often,
contact your distributor.
●
E91
E91 Memory:history
■■■■■■■■■■■■
OK(ENTER)
The memory storing the setup data is
defective.
Power was cut while writing in memory
or deleting.
●
●
If the same trouble occurs often,
contact your distributor.
As the user setting value is
initialized or is returned to the last
measured value, reset it (re-enter
it).
●
●
E92
E92 Memory:setup
■■■■■■■■■■■■
OK(ENTER)
The cable is not connected.
The setting (e.g. baud rate) of the
connected device (such as PC) is
incorrect.
●
●
Connect the cable correctly.
Confirm the setting (e.g. baud rate)
of analyzer connections (e.g. PC).
●
●
E39
E39 Communication
■■■■■■■■■■■■
OK(ENTER)
Forgot to set the Tip Waste Case.
The Tip Waste Case is not correctly set.
●
●
Set the Tip Waste Case correctly.
●
E38
E38 Tip case not set
■■■■■■■■■■■■
OK(ENTER)
When sampling, air bubbles or fibrin
was aspirated.
When sampling, the sample amount
was insufficient.
●
●
Remove air bubbles or fibrin.
Confirm the sample amount.
●
●
E37
E37 Sampling error
■■■■■■■■■■■■
OK(ENTER)
Wear protective gloves in order to avoid exposure
to pathogenic microbes.
WARNING
Failed to draw the sample and drop it to
the reagent correctly due to lack of
sample amount or absorbing fibrin.
Failed to drop the sample to the reagent
correctly because the sample adheres
around the dropping hole.
●
●
Check the sample amount and
remove fibrin if necessary.
Wipe off the sample adhered to
around the dropping hole.
Measure it with a different tip.
●
●
●
E40
E40 Sample drop
(
■■
7
■■
4
■■
1)
■■■
OK(ENTER)
(1〜9):Channels with abnormalities
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...