C
4-16 ● SP-4430
4-3
定期的なお手入れ
Chapter 4 MAINTENANCE
4-3-4
Replacement of Nozzle
With numerous measurements, the O-ring attached to the Nozzle becomes
deteriorated.
Remove the cover.
2.
Remove the top cover by sliding to
the left.
●
Requirements
Nozzle, Nozzle replacement tool and Protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
WARNING
Discard used samples, tips and protective gloves separately from general
waste according to local regulations on biohazardous waste.
CAUTION
Move the Nozzle to stand-by position.
1.
Press [2] key on the MAIN MENU.
SUB MENU 1/2 is displayed.
Press [3] key .
The Table Cover opens, and the
Reagent Table and Centrifuge-
equipped Multi Rack slide forward.
The Nozzle Driving Unit moves to
the left end in the analyzer.
Turn OFF the power.
●
●
●
1 . R e s u l t s 2 . P A R AM
3 . M a i n t e n a n c e
( 1/2 )
1 . M e a s u r e 2 . S u b m e n u
3 . C a l i b r a t e
( 1/1 )
R e a d y .
P l e a s e
t u r n o f f .
Make sure the power is OFF before moving on to the next procedure.
Otherwise, it may cause injury by contacting with the driving unit of the
analyzer.
CAUTION
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...